Series 2500 Total Hydrocarbon Analyzer
▪Methanizer Option to measure CO and CO2
▪Consistent, Precise, Flexible
▪19” Rack Mount or Bench Top
Principal of Operation
▪ Gas purity certification
▪ UHP inert gas contaminant analysis
▪ Carbon bed breakthrough detection
▪ HC monitoring in CO2 feed streams in the food/beverage industry
▪ Process monitoring
▪ Monitoring ambient air and other gases
▪ Monitoring of HC emissions
▪ R&D laboratory applications
▪ Flame Ionization Detector (FID)
▪ Methanizer to measure CO and CO2 on -CM model
▪ High resolution 7 inch graphical LCD touchscreen
▪ Flameout indicator with fail-safe gas shut-off
▪ Auto ranging across 4 ranges
▪ Auto zero and remote calibration
▪ Programmable relays for concentration alarms
▪ Electronic flow control of air, fuel, and sample gases
▪ Real-time data logging by USB and serial port
Model Model
2500 2502 Total Hydrocarbon Gas Analyzer,standard
2500-CM 2502-CM Total Hydrocarbon Gas Analyzer with catalytic methanizer
Detector
|
Flame Ionization (FID)
|
Ranges
|
User selectable nominal ranges. Upper limits are functions of signal.
Range 1: 0 - 25.00 ppm
Range 2: 0 - 250.0 ppm
Range 3: 0 - 2,500 ppm
Range 4: 0 - 25,000 ppm
|
Linear Range (+/-5%)
|
0.0034 g/sec (12,750 ppm CH4 )
|
Dynamic Range
|
0.0095 g/sec (0 to 35,560 ppm CH4 )
|
High Concentration Analysis
|
Limits are a function of lower sample flow.
Range 4: 0 – 10% CH Dynamic Range: 0 – 10% CH 4 4
Linear Range: 4% CH4
|
Sensitivity
|
Average of various samples: 3630 Asec/g
Best case CH4 in N2 : 2916 Asec/g
|
Minimum Detectability
|
Average of various samples: 0.018 ppm CH4
Best case CH4 in N2 : 0.013 ppm CH4
|
Repeatability
|
± 1% of full scale for 5 successive samples
|
Accuracy
|
99.9% per individual linear range/99.45% per full scale
|
Drif
|
Zero and span, 24 hr. constant ambient temperature, spanned at 10 ppm: < 0.2% of span
Zero and span, variable ambient temperature from 18 to 32 °C, spanned at 10 ppm: < 0.2% of span
|
Response Time
|
(Measured between 0 and 1007 ppm CH4 in N2 on range 3. Independent of inlet pressure)
Rising to 90% of final reading (sec): 16
Rising to 95% of final reading (sec): 18
Rising to 100% of final reading: < 4 min.
Falling to 10% of starting reading (sec): 16
Falling to 5% of starting reading (sec): 18
Falling to 0% of starting reading: < 6 min.
|
Communications
|
Analog: 4-20 mA, 0-5 VDC
Digital: RS-232 serial, USB data logging
|
Relays
|
Programmable relays for two (2) high/low concentration alarms
(2) 1A alarm relays and (1) ready relay
|
User Interface
|
7-inch graphical resistive LCD touchscreen
|
Gas Parameters
|
Sample: 25 ccpm typical flow to FID. Inlet pressure range 10 to 100 psig. Sample bypass flow is mini
mum at 10 psig up to about 1 lpm at 100 psig inlet pressure
Air for FID: 300 ccpm maximum flow, UHP grade (99.999%), supplied at 20 to 100 psig
Fuel for FID (Safety and minimum detectability are among considerations when selecting fuel gas
type. Fuel mix is required for oxygen sample)
100% hydrogen 40 ccpm maximum flow, UHP grade (99.999%), supplied at 20 to 100 psig
40% H2 - 60% N2 mix 100 ccpm maximum flow, UHP grade (99.999%), supplied at 20 to 100 psig
|
Gas Process Connections
|
1/8-inch compression fitting
|
Ambient Temperature
|
15 °C to 35 °C (59 °F to 95 °F) stable ambient temperature for best p
|
Power Requirement
|
2500: 120 W at 115 VAC, 60 Hz, 2500-CM: 180 W at 115 VAC, 60 Hz
2502: 120 W at 230 VAC, 50 Hz, 2502-CM: 180 W at 230 VAC, 50
|
Dimensions
|
7 in (17.8 cm) H x 19 in (48.3 cm) W x 21 in (53.4 cm) D
Front panel height 4U fits 19-inch-wide EIA standard rack
|
Weight
|
2500 or 2502: 24 pounds (11 kg) 2500-CM or 2502-CM: 27 pounds (13 kg) |